Abstract
This study compared heart rate (HR) measurements taken from two wrist-worn devices; the Empatica E4 and the Apple Watch Series 5, to that taken from a Polar H10 chest strap. Ten healthy adult volunteers took part in a laboratory validation study and performed a treadmill exercise protocol. A single-subject validity study was also conducted to evaluate the accuracy of continuous HR measurements obtained during free-living activities. The participant wore both wrist devices, as well as the Polar H10 for 12-hours, as she continued her habitual daily activities. The key findings of the laboratory study were that the Apple Watch was accurate at assessing HR compared to the Polar H10 with Mean Absolute Percentage Error (MAPE) values < 5% during treadmill exercise. The accuracy of the E4 however was generally poor with MAPE values > 15%. Findings from the single-subject validity study indicate that the Apple Watch produces accurate measurements of HR, whereas the E4 device overestimated HR, except for during the more strenuous activities undertaken where HR was underestimated. Clinical Relevance— The Apple Watch has acceptable accuracy in measuring HR during treadmill exercise and during free-living activities in healthy adult volunteers.
Citation
Giggins, O. M., Doyle, J., Sojan, N., Moran, O., Crabtree, D. R., Fraser, M., & Muggeridge, D. J. (2021, November). Accuracy of Wrist-Worn Photoplethysmography Devices at Measuring Heart Rate in the Laboratory and During Free-Living Activities. Presented at 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC, Mexico
Presentation Conference Type | Conference Paper (published) |
---|---|
Conference Name | 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC |
Start Date | Nov 1, 2021 |
End Date | Nov 5, 2021 |
Acceptance Date | Sep 30, 2021 |
Online Publication Date | Dec 9, 2021 |
Publication Date | 2021 |
Deposit Date | Dec 22, 2021 |
Publicly Available Date | Jan 10, 2022 |
Publisher | Institute of Electrical and Electronics Engineers |
Series ISSN | 2694-0604 |
Book Title | 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) |
DOI | https://doi.org/10.1109/embc46164.2021.9629522 |
Keywords | Heart rate, Wrist, Performance evaluation, Protocols, Atmospheric measurements, Measurement uncertainty, Photoplethysmography |
Public URL | http://researchrepository.napier.ac.uk/Output/2831575 |
Files
Accuracy Of Wrist-Worn Photoplethysmography Devices At Measuring Heart Rate In The Laboratory And During Free-Living Activities (accepted version)
(275 Kb)
PDF
You might also like
Clinical and health economic evaluation of a post-stroke arrhythmia monitoring service
(2022)
Journal Article
Downloadable Citations
About Edinburgh Napier Research Repository
Administrator e-mail: repository@napier.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search